เซนเซอร์แปลงความดัน Pressure Transmitter
รุ่น SMX2FG
คุณลักษณะ
The SMX2 in field housing has been designed for application in rugged environments. Thanks to its stainless-steel thin-film measuring cell, this pressure transducer features excellent properties, it isextremely robust, fully welded and without any hydraulic transmission media. The measuring range can be freely positioned at a ratio of 4:1 in the upper end range. There is also thepossibility to freely select the response time. Both parameters can be managed with the help of a service box, including software. Further advantages of this model consist in the transmitter being able to be adjusted or calibrated even without having to separate it from the measuring loop. Tests can be performed within the system by means of a loop generator (pressure simulator) inside the transmitter
Zone 0 An area where an explosive atmosphere of a mix of air and combustible gases, vapours or sprays is permanently, over long periods or frequently prevails.
Zone 1 An area where an explosive atmosphere of a mix of combustible materials in the form of gas, vapour or spray with air occurs occasionally in normal operation.
Safety Note: When fitting, commissioning and operating this pressure transmitter, please observe relevant national safety regulations by all means.
- Measuring range 0 to 1…2000 bar
- Explosion-proof certificate for zone 0 II 1G Ex ia IIB T4 Ga or II 1G Ex ia IIC T4 Ga
- Explosion-proof certificate for zone 0 II 2G Ex ia IIC T4 Gb
- Zero-function, downscale 4:1
Technische Daten | |||||||||
---|---|---|---|---|---|---|---|---|---|
Measuring range* stianless steel diaphragm | p [bar] | 1,0 | 1,6 | 2,0 | 2,5 | 4,0 | 6,0 | 10,0 | |
- Overload pressure | p [bar] | 6 | 6 | 6 | 6 | 10 | 20 | 20 | |
- Burst pressure | p [bar] | 9 | 9 | 9 | 9 | 15 | 30 | 30 | |
Measuring range* stainlees steel diaphragm | p [bar] | 16 | 20 | 25 | 40 | 60 | 100 | 160 | |
- Overload pressure | p [bar] | 40 | 40 | 100 | 100 | 200 | 200 | 400 | |
- Burst pressure | p [bar] | 60 | 60 | 150 | 150 | 300 | 300 | 600 | |
Measuring range* stainless steel diaphragm | p [bar] | 200 | 250 | 400 | 600 | 1000 | 1600 | 2000 | |
- Overload pressure | p [bar] | 400 | 750 | 750 | 840 | 1200 | 2400 | 2400 | |
- Burst pressure | p [bar] | 600 | 1000 | 1000 | 1050 | 1500 | 3000 | 3000 | |
(Vacuum, relative pressure, +- or abso;ute pressure are available), Please note:> 1000 bar with thread M18x1,5 | |||||||||
ELECTRICAL PARAMETER | |||||||||
2-wire | |||||||||
Output signal* | 4...20mA | ||||||||
Supply voltage | US [VDC] | 20...27 | |||||||
Load resistor | RA [Ω] | acc. to RA =< (US-20V)/0,02 A | |||||||
Responde time | t [ms] | ≥ 4 (digital) | |||||||
Maximum supply current | I [mA] | 23mA | |||||||
Isolation voltage | U [VDC] | 500V AC | |||||||
ACCURACY | |||||||||
pressure range 1 bar to 2000 bar | |||||||||
Genauigkeit @ RT | % of the range | ≤0,50*** | Option 0,25 | ||||||
Non-lineearity | BFSL | ≤0,15 | |||||||
Stability/year | % of the range | ≤0,15 | |||||||
*** incl. nonlinearity, hysteresis, repeatability zero-offset- and final-offset (acc. to IEC 61298-2) | |||||||||
ACCEPTABLE TEMPERATURE RANGES | |||||||||
zone 0 | zone 1 | ||||||||
Measuring medium | T [℃] | -20...60 | -40...100 | ||||||
Ambience | T [℃] | -20...60 | -40...85 | ||||||
Stroage | T [℃] | -40...120 | -40...120 | ||||||
Compensated range* | T [℃] | -20...60 | -20...85 | ||||||
Mean TC offset | % of the range | ≤ 0,15 / 10K | ≤ 0,15 / 10K | ||||||
Mean TC range | % of the range | ≤ 0,15 / 10K | ≤ 0,15 / 10K | ||||||
Total | % of the range | -20°C 1,00% | digits : -40°C 1,00% | ||||||
% of the range | 60°C 1,00% | digits : 85°C 1,00% | |||||||
Directive ATEX | |||||||||
zone 0 | zone 1 | ||||||||
Type of ignition protection | II 1G Ex ia IIB T4 Ga | II 2G Ex ia IIC T4 Gb | |||||||
Underlying standards | EN 60079-0, EN 60079-11, EN 60079-26, EN 60079-14 (both zones) | ||||||||
Maximum connected power | 27 V, 125mA, 0,85 W | 27 V, 125mA, 0,85 W | |||||||
Temperature class | T4 (Ambience -20…+60° C) | T4 (Ambience -40…+85° C) | |||||||
MECHANICAL PARAMETER | |||||||||
Parts in contact with the measuring medium* | staniless steel | for pressure range of 1 bar 2000 bar | |||||||
Housing* | Stainless steel | ||||||||
Shock resistance | g | 1000 | acc. to DIN EN 60068-2-32-free fall | ||||||
Vibration resistance | g | 20 | acc. to DIN EN 60068-2-6-vibration sinusoidal | ||||||
G-Force | g | 50 | acc. to DIN EN 60068-2-27-shock | ||||||
Mass | m [g] | ~600 | depending on design | ||||||
Approval | IBExU10ATEX1014 | ||||||||
IP system of protection (IEC 605029) up to IP69K | The IP system of protection as specified in the data sheets generally applies, with appropriate mating plug connected. | ||||||||